Physical
Power
Perfomance
Perfomance Characteristics Table
The aerosol generator featured in the Aeroneb Pro System.
| Nebulizer Unit Weight | 25 grams (0.9 oz.) |
| Control Module Weight | 230 grams (8.1 oz.) |
| Medication Cup Capacity | 10 mL |
| Consumption | 6.5 Watts (charging) 2.0 Watts (nebulizing) AC/DC adapter (input 100 to 240 to VAC 50/60Hz, output 9 V) or rechargable battery (4.8 V nominal output) |
| Patient Isolation | Approved to IEC 60601-1 and AAMI ES1 |
| Noise Level | Silent operation |
| Frequency, Nebulizer Unit | 128 kilohertz (kHz) |
| MMAD | GSD | FPF | Residual Volume | Adult % Dose Deposited | |
| Aeroneb Pro Nebulizer System | 2.1 | 2.2 | 83% | 0.3 mL | 13% |